Menu Close

Installation site preparation

Bench Space for the CAMAG DBS-MS 500

Width: 1260 mm (including handles)
Depth: 800 mm (add 250 mm at the rear for tubing and connections, add 450 mm in front to open the door)
Height: 660mm (plus solvent bottles and capillaries)

Adequate space for the waste canister underneath the instrument is necessary.
Therefore, the instrument can not be placed on the ground. The instrument must be placed at a solid desk with at least 150Kg load carrying capacity

Nitrogen/clean air supply for the CAMAG DBS-MS 500

5.5 -8 bar; 6mm tube outer diameter (If not available use clean compressed air with a water and oil trap).

LC-MS/MS communication

The start signal cable for the signal transmission from the CAMAG DBS-MS 500 to the LC-MS/MS system will be configured and supplied by CAMAG based on the hardware setup at the installation site.

Compatibility

The DBS-MS 500 is fully integrated into Shimadzu, Waters, Agilent, Sciex and Thermo LCMS instruments and software.

The coupling of a DBS-MS 500 with a Shimadzu 8060 as an example